Quality control comes up often in conversation and rarely with the context attached. Here we lay out the basics in order, then work through the practical considerations.
Last reviewed on 2025-08-08. Where a claim depends on a specific study, the study is described rather than over-claimed.
Routine HPLC testing compares a sample result with a calibration curve prepared from known reference standards. Peak area or peak height is plotted against concentration, and the curve is used to estimate unknown amounts. Retention time supports tentative identification when compared with a standard, though mass spectrometry or another confirmatory method may be needed for definitive identification. Pre-run checks verify repeatability, resolution, and peak symmetry before sample analysis. Limits of detection and quantification describe the smallest amounts that can be reliably observed or measured. Sample preparation, filtration, and degassing help prevent column damage and inconsistent results.
High-performance liquid chromatography is an analytical technique that separates components in a liquid sample. A pump moves a liquid mobile phase through a column packed with a solid stationary phase. Compounds interact differently with both phases and travel at different rates, leaving the column at distinct retention times. A detector records these arrivals as peaks on a chromatogram. The resulting pattern supports identification and quantification of substances in mixtures. Modern instruments use high pressure to force solvent through small particles, which improves speed and resolution compared with older low-pressure liquid chromatography methods.
Separation in HPLC depends on the chemistry of the stationary phase, the composition of the mobile phase, and the physical properties of the column. Reverse-phase separations use a nonpolar stationary phase and a polar mobile phase, and they are common for many organic compounds. Ion-exchange, size-exclusion, and normal-phase modes serve other classes of analytes. Gradient elution changes solvent strength over time, while isocratic elution holds it constant. Flow rate, temperature, particle size, and column length all influence peak shape and resolution. Detection may use ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry, depending on the analyte and the required sensitivity.
Routine quality control uses system suitability, blank injections, check standards, and control samples to detect drift or contamination. System suitability criteria may specify minimum resolution, maximum tailing factor, and a permitted range for repeated injections. Blank injections reveal carryover or solvent contamination, while check standards confirm calibration accuracy over a batch. Control samples with known analyte levels can show whether results remain within statistical limits. When a control result falls outside limits, the analyst investigates the cause and may invalidate affected results before repeating the batch.
Documentation and traceability are central to regulated HPLC testing. Records typically include instrument logs, column history, mobile-phase preparation, sample preparation, injection sequences, raw chromatograms, and audit trails. Electronic systems may require user access controls, time-stamped changes, and backup procedures. Training records show that analysts are qualified for assigned methods. Audits and inspections check whether written procedures match actual practice and whether deviations are documented. These controls support reproducibility and allow results to be reconstructed if questions arise later.
| Property | Value | Notes |
|---|---|---|
| Common abbreviation | HPLC | High-performance liquid chromatography |
| Separation basis | Differential partitioning | Between liquid mobile phase and solid stationary phase |
| Common mode | Reverse phase | Nonpolar column, polar mobile phase |
| Typical detector | UV-Vis absorbance | Widely used for compounds with chromophores |
| Typical column particle size | 2–5 µm | Smaller particles can improve resolution |
Method validation demonstrates that an HPLC procedure is suitable for its intended purpose. Common validation parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, and robustness. Accuracy reflects agreement with a reference value, while precision describes repeatability under defined conditions. Specificity shows whether the method can measure the analyte in the presence of impurities or matrix components. Validation documents are reviewed before a method is used for routine testing or regulatory submissions.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Typical checks include retention time, peak area precision, resolution between critical pairs, tailing factor, and theoretical plate count. Acceptance criteria are set in the method or pharmacopeial monograph. If a suitability check fails, the run may be rejected and the instrument or sample preparation may need investigation. This practice helps prevent release of data from a system that has drifted out of control.
Validation demonstrates that a method is suitable for its intended use. Typical performance characteristics include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantitation, and robustness. Regulators and standards organizations provide frameworks, but specific requirements depend on the application and jurisdiction. System suitability tests are run before sample analysis to confirm resolution, peak symmetry, retention time repeatability, and sensitivity. A validated method is not permanently fixed; changes may require partial or full revalidation.
Routine HPLC testing depends on controlled reagents, calibrated instruments, and documented procedures. Columns degrade over time, so retention times and peak shapes are monitored for drift. Mobile phases are filtered and degassed to prevent pump damage and detector noise. Reference standards must be traceable and stored under suitable conditions. Data handling systems record injections, calculations, and audit trails. Quality control samples interspersed with unknowns help detect errors during a run.
Developing an HPLC method begins with defining the purpose, such as quantifying a main component, measuring impurities, or confirming identity. Analysts select separation mode, column, mobile phase, detection, and sample preparation based on analyte properties and matrix. Experiments vary solvent strength, pH, buffer type, and temperature to achieve resolution between critical peaks. The goal is a robust method that produces reliable results across instruments and operators. Method development often involves trial runs and statistical optimization.
Accelerator mass spectrometry (AMS) is a form of mass spectrometry that accelerates ions to extraordinarily high kinetic energies before mass analysis. The special strength of AMS among the different methods of mass spectrometry is its ability to separate a rare isotope from an abundant neighboring mass ("abundance sensitivity", e.g. 14C from 12C). The method suppresses molecular isobars completely and in many cases can also separate atomic isobars (e.g. 14N from 14C). This makes possible the detection of naturally occurring, long-lived radio-isotopes such as 10Be, 36Cl, 26Al and 14C. (Their typical isotopic abundance ranges from 10−12 to 10−18.) AMS can outperform the competing technique of decay counting for all isotopes where the half-life is long enough. Other advantages of AMS include its short measuring time as well as its ability to detect atoms in extremely small samples.
Automated iterative homologation enables stepwise construction of carbon chains through repeated one-carbon extensions of boronic esters. Two reactions adapted to automation are the Matteson homologation, which inserts a methylene unit via chloromethyllithium, and chiral carbenoid homologation, which uses lithiated benzoate esters. Both methods have been implemented on robotic platforms under low-temperature, inert conditions. Using these techniques, up to six consecutive C(sp³)–C(sp³) bond-forming homologations have been performed without manual intervention, representing the highest number reported in an automated synthesis. The approach has also been applied to the synthesis of intermediates of the natural product (+)-kalkitoxin.
Archives of Biochemistry and Biophysics is a biweekly peer-reviewed scientific journal that covers research in biochemistry and biophysics. It is published by Elsevier. As of 2012, the editors-in-chief were Paul Fitzpatrick (University of Texas Health Science Center at San Antonio), Helmut Sies (University of Düsseldorf), Jian-Ping Jin (Wayne State University School of Medicine), and Henry Jay Forman (University of Southern California). The journal was established in 1942 by Academic Press as the Archives of Biochemistry, obtaining its current name in 1952. It absorbed the journal Molecular Cell Biology Research Communications (formerly section B of Biochemical and Biophysical Research Communications), which was published from 1999 to 2001. An index to authors for the first 75 volumes, covering the period from 1943 to 1958, was published in October 1959.
The adipokines, or adipocytokines (Greek adipo-, fat; cytos-, cell; and -kinos, movement) are cytokines (cell signaling proteins) secreted by adipose tissue. Some contribute to an obesity-related low-grade state of inflammation or to the development of metabolic syndrome, a constellation of diseases including, but not limited to, type 2 diabetes, cardiovascular disease and atherosclerosis. The first adipokine to be discovered was leptin in 1994. Since that time, hundreds of adipokines have been discovered. Members include: Leptin Adiponectin Apelin chemerin interleukin-6 (IL-6) monocyte chemotactic protein-1 (MCP-1) plasminogen activator inhibitor-1 (PAI-1) retinol binding protein 4 (RBP4) tumor necrosis factor visfatin omentin vaspin (SERPINA12) progranulin CTRP-4 Interleukin 8 (IL-8), interleukin 10 (IL-10), interferon gamma (IFN-γ) and inducible protein 10 (IP-10 or CXCL10) have been shown to be associated with excessive body weight. Adipose tissue Hepatokines Myokines Adipokines at the U.S. National Library of Medicine Medical Subject Headings (MeSH)
Sources: en.wikipedia.org
Fermentation is the metabolism of organic compounds in the absence of air. It involves substrate-level phosphorylation in the absence of a respiratory electron transport chain. The equation for the reaction of glucose to form lactic acid is: C6H12O6 + 2 ADP + 2 Pi → 2 CH3CH(OH)COOH + 2 ATP + 2 H2O Anaerobic respiration is respiration in the absence of O2. Prokaryotes can utilize a variety of electron acceptors. These include nitrate, sulfate, and carbon dioxide. In anaerobic organisms and prokaryotes, different pathways result in ATP. ATP is produced in the chloroplasts of green plants in a process similar to oxidative phosphorylation, called photophosphorylation. ATP can also be synthesized through several so-called "replenishment" reactions catalyzed by the enzyme families of nucleoside diphosphate kinases (NDKs), which use other nucleoside triphosphates as a high-energy phosphate donor, and the ATP:guanido-phosphotransferase family.
Viral induction of apoptosis occurs when one or several cells of a living organism are infected with a virus, leading to cell death. Cell death in organisms is necessary for the normal development of cells and the cell cycle maturation. It is also important in maintaining the regular functions and activities of cells. Viruses can trigger apoptosis of infected cells via a range of mechanisms including:
The Association of Public Health Laboratories (APHL) is a membership organization in the United States representing the laboratories that protect the health and safety of the public. APHL serves as a liaison between public health laboratories and federal and international agencies. Membership consists of local, state, county, and territorial public health laboratories; public health environmental, agricultural and veterinary laboratories; and corporations and individuals with an interest in public health and laboratory science. APHL is a non-profit, 501(c)(3) organization with a history of over fifty years.
Sources: en.wikipedia.org
HPLC testing measures the presence and amount of one or more compounds in a liquid sample. It separates mixture components and records detector responses as peaks, which are compared with reference standards. Results are usually reported as concentrations or relative percentages.
Retention time is the interval between sample injection and the detector response for a given compound. It depends on the compound's interactions with the stationary and mobile phases under set conditions. Matching a retention time to a standard supports tentative identification but is not always unique.
HPLC alone can separate unknown compounds and provide retention times, but it often cannot identify them with certainty. Coupling HPLC to mass spectrometry gives mass information that improves identification. Confirmation usually requires comparison with reference standards or complementary techniques.
Validation establishes suitability for a new method, while verification confirms that a method works in a specific laboratory. Verification is often used when a validated method is adopted with existing equipment and staff. Both rely on documented acceptance criteria.